Corneal thickness measurements with the Orbscan Topography System and ultrasonic pachymetry
- 1 November 1997
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 23 (9) , 1345-1350
- https://doi.org/10.1016/s0886-3350(97)80113-7
Abstract
Purpose: To compare corneal thickness measurements obtained with a new instrument, the Orbscan Topography System, with those obtained with the DGH ultrasonic pachymeter and to assess the agreement and repeatability of the two devices. Setting: LSU Eye Center, New Orleans, Louisiana, USA. Methods: Measurement agreement was assessed in 51 eyes of 26 normal volunteers using both Orbscan and ultrasonic pachymetry. Repeatability for the instruments was measured in 10 eyes of 5 additional volunteers. Corneal thicknesses were compared using the analysis of variance (ANOVA). The relationship between the devices was assessed by analysis of regression (ANOR). Results: In the measurement agreement experiment, the mean corneal thickness was 571.3 μm ± 6.21 SEM with the Orbscan system and 543.3 ± 7.49 μm with ultrasonic pachymetry; these values were significantly different (F test, ANOVA, P = .0048). In the repeatability experiment, the mean thickness was 561.1 ± 8.42 μm with the Orbscan system and 537.4 ± 5.84 μm with ultrasound pachymetry; these values were also significantly different (F test, ANOVA, P = .0003). Analysis of regression showed a significant linear regression between the values obtained with the devices (P = .0001, F test, ANOR). Conclusions: In both studies, the Orbscan system obtained statistically significantly different and higher values for corneal thickness. Regression analysis suggests that over the range of values in this study, the two devices differ by a constant amount (intercept and slope). The nonzero intercept of this regression shows that the values from the devices differ and cannot be directly substituted for each other. We therefore conclude that in this study, Orbscan system measurements of corneal thickness were 23 to 28 μm greater than ultrasonic pachymeter measurements. Linear regression equations may be developed for the results of measurements from the two devices and used as a precise transformation factor for the values obtained with the two devices.Keywords
This publication has 10 references indexed in Scilit:
- Reliability Coefficients of Three Corneal PachymetersAmerican Journal of Ophthalmology, 1992
- Comparison of Central Corneal Thickness Measurements between Optical and Ultrasound PachometersOptometry and Vision Science, 1992
- Variability of Ultrasonic PachometryOptometry and Vision Science, 1990
- Eye-length measurement by interferometry with partially coherent lightOptics Letters, 1988
- Differences Among Ultrasonic Pachymeters in Measuring Corneal ThicknessJournal of Refractive Surgery, 1987
- Some components affecting the precision of corneal thickness measurement performed by optical pachometryActa Ophthalmologica, 1986
- Method of Evaluating Ultrasonic PachometersOptometry and Vision Science, 1986
- Comparison of Ultrasonic Corneal Thickness Measurements Before and During Surgery in the Prospective Evaluation of Radial Keratomy (PERK) StudyOphthalmology, 1986
- Femtosecond optical ranging in biological systemsOptics Letters, 1986