Experimental analysis of T1 imaging with a single‐scan, multiple‐point, inversion‐recovery technique
- 1 June 1992
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 25 (2) , 337-343
- https://doi.org/10.1002/mrm.1910250212
Abstract
Look and Locker's single‐scan, multiple‐point, T1 determination technique modified for imaging applications was evaluated experimentally for its accuracy and reliability with respect to the pulse sequence parameters. in particular. to the interpulse delay, TD, and the tip angle, α. T1 imaging experiments were performed with a 0.5‐T imaging system on a phantom which consisted of an array of vials containing 2% agarose gels doped with various amount of Mn2+ using different combinations of the parameter set {TD. α}. T1, results obtained with this technique were compared with those measured by a conventional inversion‐recovery procedure using the same spectrometer. Strategic choice of pulse sequence parameters to minimize experimental errors and the criteria for these choices will be discussed. © 1992 Academic Press, Inc.Keywords
This publication has 6 references indexed in Scilit:
- A comparison of one‐shot and recovery methods in T1 imagingMagnetic Resonance in Medicine, 1988
- Tip-angle-reduced T1 imagingJournal of Magnetic Resonance (1969), 1986
- NMR imaging of spin-lattice relaxation using stimulated echoesJournal of Magnetic Resonance (1969), 1985
- Strategies and tactics in NMR imaging relaxation time measurements. I. Minimizing relaxation time errors due to image noise—the ideal caseMagnetic Resonance in Medicine, 1985
- The choice of optimal parameters for measurement of spin-lattice relaxation times. II. Comparison of saturation recovery, inversion recovery, and fast inversion recovery experimentsJournal of Magnetic Resonance (1969), 1980
- Time Saving in Measurement of NMR and EPR Relaxation TimesReview of Scientific Instruments, 1970