Examination of the performance exhibited by a single detector double focusing magnetic sector ICP-MS instrument for uranium isotope abundance ratio measurements over almost three orders of magnitude and down to pg g−1 concentration levels
- 3 February 2000
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 15 (4) , 353-358
- https://doi.org/10.1039/a907084b
Abstract
Calibration experiments were performed on a commercially available double focusing magnetic sector single detector ICP-MS instrument to evaluate its potential for accurate isotope abundance ratio measurements down to the 1 pg g−1 isotope concentration level (without involving any preconcentration step). Its performance, for isotope abundance ratios varying over almost three orders of magnitude from ≈1∶1 to ≈1∶500, was checked using the IRMM-072 series of Isotopic Reference Materials. Routine experimental settings were applied and remained unchanged for the entire series of experiments. It is shown that all the certified ratios (certified to within 0.03% expanded combined uncertainty, k = 2) could be matched experimentally within the expanded (k = 2) combined uncertainty of the measurements (from ±0.11 to ±0.33% for 233U/235U, and from ±0.16 to ±0.31% for 233U/238U). For 1∶1 ratios (235U/238U) a 0.05% precision could be reproducibly obtained of which only 33% could be accounted for by Poisson statistics, whereas for 1∶500 ratios a 0.16% precision was obtained which could be completely explained by the same Poisson theory. In addition to the experimental precision other possible sources of uncertainty were investigated. The way isotope abundance ratios are measured (acceleration voltage is varied while the magnetic field is kept constant) is discussed as a potential extra source of mass discrimination. The crucial importance of a proper correction for dead-time effects is demonstrated. Finally, the hypothesis of a contamination with natural uranium that would have biased the results could be ruled out for all samples.Keywords
This publication has 21 references indexed in Scilit:
- A new variable dispersion double-focusing plasma mass spectrometer with performance illustrated for Pb isotopesInternational Journal of Mass Spectrometry, 1998
- High resolution inductively coupled plasma mass spectrometry for the trace determination of plutonium isotopes and isotope ratios in environmental samplesJournal of Analytical Atomic Spectrometry, 1998
- Strontium isotope ratio measurements in prehistoric human bone samples by means of high-resolution inductively coupled plasma mass spectrometry (HR-ICP-MS)Journal of Analytical Atomic Spectrometry, 1998
- Isotopic and ultratrace analysis of uranium by double-focusing sector field ICP mass spectrometryAnalytical and Bioanalytical Chemistry, 1997
- Determination of lead concentrations and isotope ratios in recent snow samples from high alpine sites with a double focusing ICP-MSAnalytical and Bioanalytical Chemistry, 1997
- Low Blank Preconcentration Technique for the Determination of Lead, Copper, and Cadmium in Small-Volume Seawater Samples by Isotope Dilution ICPMSAnalytical Chemistry, 1997
- Direct Determination of Heavy Metals at Picogram per Gram Levels in Greenland and Antarctic Snow by Double Focusing Inductively Coupled Plasma Mass SpectrometryJournal of Analytical Atomic Spectrometry, 1997
- Applicability of High-Resolution ICP−Mass Spectrometry for Isotope Ratio MeasurementsAnalytical Chemistry, 1997
- Precise Measurement of Isotope Ratios with a Double-Focusing Magnetic Sector ICP Mass SpectrometerAnalytical Chemistry, 1996
- Communication. Inductively coupled plasma as an ion source for high-resolution mass spectrometryJournal of Analytical Atomic Spectrometry, 1989