Comparison of the GenMark Diagnostics eSensor Respiratory Viral Panel to Real-Time PCR for Detection of Respiratory Viruses in Children
- 1 November 2012
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 50 (11) , 3458-3465
- https://doi.org/10.1128/jcm.01384-12
Abstract
A novel eSensor respiratory viral panel (eSensor RVP) multiplexed nucleic acid amplification test (GenMark Diagnostics, Inc., Carlsbad, CA) was compared to laboratory-developed real-time PCR assays for the detection of various respiratory viruses. A total of 250 frozen archived pediatric respiratory specimens previously characterized as either negative or positive for one or more viruses by real-time PCR were examined using the eSensor RVP. Overall agreement between the eSensor RVP and corresponding real-time PCR assays for shared analytes was 99.2% (kappa = 0.96 [95% confidence interval {CI}, 0.94 to 0.98]). The combined positive percent agreement was 95.4% (95% CI, 92.5 to 97.3); the negative percent agreement was 99.7% (95% CI, 99.4 to 99.8). The mean real-time PCR threshold cycle (CT) value for specimens with discordant results was 39.73 (95% CI, 38.03 to 41.43). Detection of coinfections and correct identification of influenza A virus subtypes were comparable between methods. Of note, the eSensor RVP rhinovirus assay was found to be more sensitive and specific than the corresponding rhinovirus real-time PCR. In contrast, the eSensor RVP adenovirus B, C, and E assays demonstrated some cross-reactivity when tested against known adenovirus serotypes representing groups A through F. The eSensor RVP is robust and relatively easy to perform, it involves a unique biosensor technology for target detection, and its multiplexed design allows for efficient and simultaneous interrogation of a single specimen for multiple viruses. Potential drawbacks include a slower turnaround time and the need to manipulate amplified product during the protocol, increasing the possibility of contamination.Keywords
This publication has 40 references indexed in Scilit:
- Comparison of the Idaho Technology FilmArray System to Real-Time PCR for Detection of Respiratory Pathogens in ChildrenJournal of Clinical Microbiology, 2012
- Multicenter Evaluation of the BD Max GBS Assay for Detection of Group B Streptococci in Prenatal Vaginal and Rectal Screening Swab Specimens from Pregnant WomenJournal of Clinical Microbiology, 2010
- Nano/Microfluidics for diagnosis of infectious diseases in developing countriesAdvanced Drug Delivery Reviews, 2009
- Cost Analysis of Multiplex PCR Testing for Diagnosing Respiratory Virus InfectionsJournal of Clinical Microbiology, 2009
- Detection of Respiratory Viruses by Molecular MethodsClinical Microbiology Reviews, 2008
- Comparison of the Luminex xTAG Respiratory Viral Panel with In-House Nucleic Acid Amplification Tests for Diagnosis of Respiratory Virus InfectionsJournal of Clinical Microbiology, 2008
- Genotype Prevalence and Risk Factors for Severe Clinical Adenovirus Infection, United States 2004-2006Clinical Infectious Diseases, 2007
- Development of a Respiratory Virus Panel Test for Detection of Twenty Human Respiratory Viruses by Use of Multiplex PCR and a Fluid Microbead-Based AssayJournal of Clinical Microbiology, 2007
- A Comparative Study of Five Technologically Diverse CFTR Testing PlatformsThe Journal of Molecular Diagnostics, 2007
- Comparison of Real-Time PCR Assays with Fluorescent-Antibody Assays for Diagnosis of Respiratory Virus Infections in ChildrenJournal of Clinical Microbiology, 2006