The Finapres 2300e finger cuff
Open Access
- 1 July 1993
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 48 (7) , 611-615
- https://doi.org/10.1111/j.1365-2044.1993.tb07129.x
Abstract
Blood pressure measurements associated with malapplication of a finger cuff were compared with contemporaneous intra‐arterial pressure data in seven volunteers to determine the influence of cuff application on the accuracy of the Finapres 2300e. Systolic readings in all cuff positions differed from arterial line data by more than the recommended standard and tight and loose cuff applications under and over‐read respectively, in all subjects. The results show the Finapres to be sensitive to small degrees of finger cuff malapplication which contribute to the bias on direct arterial comparison and limit the reliability of the instrument in clinical practice.Keywords
This publication has 14 references indexed in Scilit:
- The upgraded Finapres 2300eAnaesthesia, 1992
- Continuous Noninvasive Finger Blood Pressure during Controlled Hypotension A Comparison with Intraarterial PressureAnesthesiology, 1991
- Simultaneous Comparison of Intraarterial, Oscillometric, and Finapres Monitoring During AnesthesiaAnesthesia & Analgesia, 1991
- COMPARISON OF THE FINAPRES AND DIRECT ARTERIAL PRESSURE MONITORING DURING PROFOUND HYPOTENSIVE ANAESTHESIABritish Journal of Anaesthesia, 1991
- COMPARISON OF INVASIVE AND NON-INVASIVE MEASUREMENT OF CONTINUOUS ARTERIAL PRESSURE USING THE FINAPRESBritish Journal of Anaesthesia, 1991
- The relative accuracies of two automated noninvasive arterial pressure measurement devicesJournal of Clinical Monitoring and Computing, 1991
- Noninvasive continuous blood pressure measurement from the finger: Optimal measurement conditions and factors affecting reliabilityJournal of Clinical Monitoring and Computing, 1987
- Direct Blood Pressure Measurement —Dynamic Response RequirementsAnesthesiology, 1981