Integrated optic sensor with macro-flow cell

Abstract
An integrated-optic channel waveguide device is configured as a biosensor. The device measures a refractive index change on the waveguide surface, so it is called a biorefractometer. With an appropriate overlay or selective coating, the device can monitor proteins in blood or pollutants and bio-warfare agents in water. We describe the design, fabrication, and testing of a sensor employing a waveguide Mach-Zehnder interferometer configuration. The device is fabricated in a glass substrate using potassium ion exchange. A patterned glass buffer layer defines the sensing and reference arms of the interferometer. A silicone-rubber macro-flow cell confines the liquid above the integrated-optical waveguide device. Salt solution data show that the biorefractometer has a sensitivity ((Delta) neff/(Delta) nLiquid) of 2 X 10-3 and can measure refractive index changes of about 0.005. Data obtained for antigen-antibody binding of the protein IgG indicate that a 10 percent signal change occurs in approximately 1 minute for a 10 (mu) g/ml concentration level.