Large aperture cube corner interferometer with a resolution of 0001 cm^−1
- 20 June 1991
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 30 (18) , 2575-2578
- https://doi.org/10.1364/ao.30.002575
Abstract
The interferometer of the Fourier transform spectrometer at the University of Oulu has been modified so that the maximum instrumental resolution is better than 10−3 cm−1. The resolution of the previous interferometer was 4.5 × 10−3 cm−1. The present interferometer consists of large cube corner mirrors and a large Mylar beam splitter. Each corner mirror has been made with three flat mirrors on an adjustable supporting frame. The interferometer was already in practical use in 1985. The first spectra (H2O, CO2, N2O, OCS) recorded on this interferometer have been presented in HANDBOOK OF INFRARED STANDARDS WITH SPECTRAL MAPS AND TRANSITION ASSIGNMENTS BETWEEN 3 AND 2600 μm, G. Guelachvili and K. Narahari Rao, Eds. (Academic, New York, 1986).Keywords
This publication has 3 references indexed in Scilit:
- Working resolution of 0010 cm^−1 between 20 cm^−1 and 1200 cm^−1 by a Fourier spectrometerApplied Optics, 1979
- Double-beam high resolution Fourier spectrometer for the far infraredApplied Optics, 1975
- Some More Aspects of the Michelson Interferometer with Cube CornersJournal of the Optical Society of America, 1960