Automation of photochemical kinetic determinations without irradiation of thedetection cell
Open Access
- 1 January 1991
- journal article
- research article
- Published by Wiley in Journal of Analytical Methods in Chemistry
- Vol. 13 (3) , 111-113
- https://doi.org/10.1155/s1463924691000214
Abstract
This paper reports on the development of an open-closed configuration for continuous kinetic methods based on photochemical reactions; the configuration removes the need to use special flow cells for simultaneous irradiation and detection. The sample plug was recirculated through a circuit including the detector and a coil irradiated with an ultra-violet source. A multipeak recording per injected sample was obtained which allowed signal increments between peaks (whether successive or not) to be measured. The configuration was applied to the fluorimetric determination of thioridazine. The log-log calibration curve obtained was linear in the range over 0.1-20.0 μg/ml of the analyte. The sampling frequencies achieved were between 24 and 13 h-1, depending on the peaks which were selected for measurement. The method was applied to the determination of the analyte in pharmaceuticals.Keywords
Funding Information
- Consejo Nacional de Ciencia y Tecnología (PA86-0146)
This publication has 0 references indexed in Scilit: