The Hot-Film Anemometer – a Method for Blood Velocity Determination
- 1 January 1980
- journal article
- research article
- Published by S. Karger AG in European Surgical Research
- Vol. 12 (2) , 140-148
- https://doi.org/10.1159/000128119
Abstract
In an in vitro flowmodel a constant temperature hot-film anemometer was tested, using conical needle and catheter-mounted probes. Calibration, linearization and zero-point determination as well as sensitivity change with hematocrit, angulation, flow direction and dirt deposits on the film are described. Curves were compared with those obtained simultaneously from an electromagnetic flowmeter. The hot-film anemometer was direction-insensitive and signals were seen 0.01—0.03 sec before the flowmeter signals. The frequency response of the anemometer was sufficiently high to register turbulent flow. In 27 simultaneous measurements the mean anemometer results were 6 ± 8% ( ± SD) higher than the flowmeter results and the peak results correspondingly 16 ± 6%. Both differences were significant (p < 0.001). Qualitatively hot-film anemometer curves were comparable to electromagnetic flowmeter curves. However, certain differences were demonstrated.Keywords
This publication has 5 references indexed in Scilit:
- Turbulent blood flow in the ascending aorta of humans with normal and diseased aortic valves.Circulation Research, 1976
- Measurements of disordered flows distal to subtotal vascular stenoses in the thoracic aortas of dogs.Circulation Research, 1976
- Turbulent blood flow in humans: its primary role in the production of ejection murmurs.Circulation Research, 1976
- Hot film coronary artery velocity measurements in horsesCardiovascular Research, 1976
- THE VELOCITY OF BLOOD FLOW IN ARTERIES IN ANIMALSAmerican Journal of Physiology-Legacy Content, 1936