Dose enhancement from backscattered radiation at tissue–metal interfaces irradiated with high energy electrons
- 1 May 1980
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 53 (629) , 466-470
- https://doi.org/10.1259/0007-1285-53-629-466
Abstract
Irradiation of a tissue-metal interface with 13-20 MeV electrons results in an increased dose to the tissue on the entrance side of the metal. Ionization measurements were made with a thin-window parallel-plate chamber to determine the magnitude of the dose enhancement as a function of incident electron energy, thickness and atomic number of metals introduced into the electron beam. The presence of a metal resulted in a dose ranging from 6% to .apprx. 50% greater than that measured with no metal in the beam. Most of this increase in dose may be eliminated by the addition of 1-2 g/cm2 of low Z [atomic number] material between tissue and metal. [Electron beams are frequently used in the treatment [in humans]].This publication has 5 references indexed in Scilit:
- Field shaping in electron beam therapyThe British Journal of Radiology, 1976
- Physical measurements and clinical observations on the backscatter of 10MeV electrons from lead shieldingThe British Journal of Radiology, 1975
- Back-scattering from metals in superficial therapy with high energy electronsThe British Journal of Radiology, 1974
- Modification of Dose Distribution in High-Energy Electron Beam TreatmentRadiology, 1971
- Effect of Chamber Voltage on Electron Build-up MeasurementsRadiology, 1954