Corrosion of surface defects in fine wires
- 1 November 1975
- journal article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 9 (6) , 597-610
- https://doi.org/10.1002/jbm.820090606
Abstract
Defects were observed on the surfaces of various fine diameter wires commonly used in biomedical applications. These surface irregularities were viewed at high magnifications using a scanning electron microscope which has a much greater depth of field than normal light microscopy. Defects include scratches, pits, and crevices, which are the result of commercial wire drawing practices. Corrosion test results show that imperfections can serve as sites for localized corrosion attack which could lead to premature failures.Keywords
This publication has 8 references indexed in Scilit:
- The delamination theory of wearWear, 1973
- Cathodic Protection to Prevent Crevice Corrosion of Stainless Steels in Halide MediaCorrosion, 1973
- Corrosion behaviour of surgical implant materials: II. Effects of surface preparationCorrosion Science, 1969
- Electrochemical Detection of Localized CorrosionCorrosion, 1969
- Corrosion and fracture of type 316 SMO orthopedic implantsJournal of Biomedical Materials Research, 1969
- General and Intergranular Corrosion of Austenitic Stainless Steels in AcidsJournal of the Electrochemical Society, 1959
- Analysis of Corrosion Pitting by Extreme-Value Statistics And Its Application to Oil Well Tubing Caliper Surveys★Corrosion, 1957
- The “probability” of corrosionTransactions of the Faraday Society, 1935