This paper describes the design and construction of a biomechanical model that will determine the pull-out strength and fatigue failure of internal fixation devices used at the lumbosacral junction. Fusions at this junction have the highest failure rate. It is suggested that devices that successfully fix this vertebral unit will be effective at other levels of the spine. A model that tests any device designed for use at the lumbosacral junction will therefore provide a uniform measure of the effectiveness of different kinds of instrumentation at the L5–S1 vertebral unit. It will also provide predictive values for fixation. systems at other spinal levels. A real-time data-acquisition system was also designed and used to define and determine the failure of the Luque orthopaedic bone-plate fixation device.Keywords
This publication has 5 references indexed in Scilit:
- Biomechanical Comparison of Seven Internal Fixation Devices for the Lumbosacral JunctionSpine, 1987
- Internal Fixation in Lumbosacral Spine Fusion A Biomechanical and Clinical StudyPublished by Wolters Kluwer Health ,1986
- Comparison of Lumbosacral Fixation DevicesPublished by Wolters Kluwer Health ,1986
- A Biomechanical Study of Intrapeduncular Screw Fixation in the Lumbosacral SpineClinical Orthopaedics and Related Research, 1986
- Anthropometric Studies of the Human Sacrum Relating to Dorsal Transsacral Implant DesignsClinical Orthopaedics and Related Research, 1986