A High Pressure Differential Manometer
- 1 May 1963
- journal article
- Published by ASME International in Journal of Heat Transfer
- Vol. 85 (2) , 180-181
- https://doi.org/10.1115/1.3686049
Abstract
A unique manometer has been designed and successfully used to measure pressure drops of a few millimeters of mercury at static pressures up to 10,000 psia. The manometer, which utilizes a column of water and mercury, is a primary standard because the mercury surface is visible at all times. It is designed for static pressures up to 20,000 psi; however, very minor revisions would make the manometer suitable for pressures up to 50,000 psi. A description of the manometer and pertinent equations is included.Keywords
This publication has 0 references indexed in Scilit: