Array Biosensor for Simultaneous Identification of Bacterial, Viral, and Protein Analytes
- 24 July 1999
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 71 (17) , 3846-3852
- https://doi.org/10.1021/ac981425v
Abstract
The array biosensor was fabricated to analyze multiple samples simultaneously for multiple analytes. The sensor utilized a standard sandwich immunoassay format: Antigen-specific “capture” antibodies were immobilized in a patterned array on the surface of a planar waveguide and bound analyte was subsequently detected using fluorescent tracer antibodies. This study describes the analysis of 126 blind samples for the presence of three distinct classes of analytes. To address potential complications arising from using a mixture of tracer antibodies in the multianalyte assay, three single-analyte assays were run in parallel with a multianalyte assay. Mixtures of analytes were also assayed to demonstrate the sensor's ability to detect more than a single species at a time. The array sensor was capable of detecting viral, bacterial, and protein analytes using a facile 14-min assay with sensitivity levels approaching those of standard ELISA methods. Limits of detection for Bacillus globigii, MS2 bacteriophage, and staphylococcal enterotoxin B (SEB) were 105 cfu/mL, 107 pfu/mL, and 10 ng/mL, respectively. The array biosensor also analyzed multiple samples simultaneously and detected mixtures of the different types of analytes in the multianalyte format.Keywords
This publication has 18 references indexed in Scilit:
- A fiber optic biosensor for multianalyte detection: importance of preventing fluorophore aggregationSensors and Actuators B: Chemical, 1998
- Step-and-Repeat Photopatterning of Protein Features Using Caged-Biotin−BSA: Characterization and ResolutionLangmuir, 1998
- Optical immunoprobe development for multiresidue monitoring in waterAnalytica Chimica Acta, 1998
- Multianalyte sensor for the simultaneous determination of glucose, L-lactate and uric acid based on a microelectrode arraySensors and Actuators B: Chemical, 1997
- Patterned planar array immunosensor for multianalyte detectionJournal of Biomedical Optics, 1997
- MastheadAnalytical Chemistry, 1996
- Simultaneous determination of follicle stimulating hormone and luteinising hormone using a multianalyte immunosensorAnalytica Chimica Acta, 1995
- Fluorescence spectroscopy and its application to a new generation of high sensitivity, multi-microspot, multianalyte, immunoassayClinica Chimica Acta; International Journal of Clinical Chemistry, 1990
- The three-dimensional structure of the bacterial virus MS2Nature, 1990
- Complete nucleotide sequence of bacteriophage MS2 RNA: primary and secondary structure of the replicase geneNature, 1976