Toward a final design for the Birmingham boron neutron capture therapy neutron beam

Abstract
This paper is concerned with the proposed Birmingham accelerator-based epithermal neutron beam for boron neutron capture therapy (BNCT). Details of the final moderator design, such as beam delimiter, shield, and beam exit surface shape are considered. Monte Carlo radiation transport simulations with a head and body phantom have shown that a simple flat moderator beam exit surface is preferable to the previously envisioned spherical design. Dose rates to individual body organs during treatment have been calculated using a standard MIRD phantom. We have shown that a simple polyethylene shield, doped with natural lithium, is sufficient to provide adequate protection to the rest of the body during head irradiations. The effect upon the head phantom dose distributions of the use of such a shield to delimit the therapy beam has been evaluated.