A Modified Post Processing Correction Matrix For SPECT

Abstract
A post reconstruction method of attenuation compensation for Single Photon Emission Computed Tomography (SPECT) has been investigated that offers a new approach to the problem of quantitation. A modified correction matrix is generated for attenuation compensation in which the Linear Attenuation Coefficient (LAC) for each pixel is assigned a value depending on the radial distance of the pixel from the true section boundary. Attenuation compensation of transverse section images of small and large volume sources of Tc-99m in phantoms using this modified matrix indicated that a known quantity of radionuclide could be determined to better than 10%. The scatter fraction was estimated as the difference in the corrected section images using a multiplicative matrix generated with a constant LAC for each pixel and the modified matrix proposed in this report.© (1986) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.