Electron Monochromator Design

Abstract
A study has been made of all the known factors which limit the performance of high resolution (0.07 to 0.01 eV FWHM) monochromators. These limiting factors have been incorporated into design equations for the optimum (maximum current output) monochromator. The conclusions are tested by performance measurements on a prototype instrument. The results require the introduction into the design equation of a new limiting factor, an anomalous energy spread in dense electron beams, which is empirically determined.

This publication has 18 references indexed in Scilit: