Abstract
We demonstrate a rapid and highly sensitive all-electronic technique based on the resistive pulse method of particle sizing with a pore to detect the binding of unlabeled antibodies to the surface of latex colloids. Here, we use an on-chip pore to sense colloids derivatized with streptavidin and measure accurately their diameter increase on specific binding to several different types of antibodies. We show the sensitivity of this technique to the concentration of free antibody and that it can be used to perform immunoassays in both inhibition and sandwich configurations. Overall, our technique does not require labeling of the reactants and is performed rapidly by using very little solution, and the pore itself is fabricated quickly and inexpensively by using soft lithography. Finally, because this method relies only on the volume of bound ligand, it can be generally applied to detecting a wide range of ligand-receptor binding reactions.Keywords
This publication has 33 references indexed in Scilit:
- Immunosensors—principles and applications to clinical chemistryClinica Chimica Acta; International Journal of Clinical Chemistry, 2001
- Developments in Immunoassay TechnologyMethods, 2000
- Surface Plasmon Resonance-Based ImmunoassaysMethods, 2000
- Counting polymers moving through a single ion channelNature, 1994
- Off-axis response for particles passing through long apertures in Coulter-type countersMeasurement Science and Technology, 1990
- Experimental measurements of the temporal evolution of cluster size distributions for high-functionality antigens cross-linked by antibodyMacromolecules, 1983
- Measurement of the Cluster Size Distributions for High Functionality Antigens Cross-Linked by AntibodyMacromolecules, 1980
- A rapid and sensitive method for the quantitation of microgram quantities of protein utilizing the principle of protein-dye bindingAnalytical Biochemistry, 1976
- The Relationship between Particle Size and Signal in Coulter-Type CountersReview of Scientific Instruments, 1971
- Books and Publicatons ReceivedInternational & Comparative Law Quarterly, 1953