Examination of Two Issues concerning Functional Gain Measurements
- 1 February 1987
- journal article
- research article
- Published by American Speech Language Hearing Association in Journal of Speech and Hearing Disorders
- Vol. 52 (1) , 56-63
- https://doi.org/10.1044/jshd.5201.56
Abstract
The test-retest variability of aided sound-field thresholds and the feasibility of using corrected 2-cm3 coupler gain values to predict individual functional gain were examined. Test-retest data were used to generate critical differences (in dB) for statistical significance between two aided sound-field thresholds. To be significantly different at the .05 level, two aided thresholds would have to differ by greater than 15 dB. Functional gain and 2-cm3 coupler gain were compared for 20 subjects using careful measurement procedures. Individual variability in the difference between functional gain and 2-cm3 coupler gain was reduced substantially from previous studies and indicates that corrected coupler gain measurements may have some clinical utility.This publication has 5 references indexed in Scilit:
- Probe-Determined Hearing-Aid Gain Compared to Functional and Coupler GainsJournal of Speech, Language, and Hearing Research, 1985
- Test of the Assumptions Underlying Comparative Hearing Aid EvaluationsJournal of Speech and Hearing Disorders, 1983
- A Comparison of Functional Gain and 2 CM 3 Coupler GainJournal of Speech and Hearing Disorders, 1982
- Speech-Discrimination Scores Modeled as a Binomial VariableJournal of Speech and Hearing Research, 1978
- The Reliability and Validity of the Comfort Level Method of Setting Hearing Aid GainJournal of Speech and Hearing Disorders, 1977