Bioabsorbable polymeric material coils for embolization of intracranial aneurysms: a preliminary experimental study
- 1 March 2001
- journal article
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of Neurosurgery
- Vol. 94 (3) , 454-463
- https://doi.org/10.3171/jns.2001.94.3.0454
Abstract
Object. A new embolic agent, bioabsorbable polymeric material (BPM), was incorporated into Guglielmi detachable coils (GDCs) to improve long-term anatomical results in the endovascular treatment of intracranial aneurysms. The authors investigated whether BPM-mounted GDCs (BPM/GDCs) accelerated the histopathological transformation of unorganized blood clot into fibrous connective tissue in experimental aneurysms created in swine. Methods. Twenty-four experimental aneurysms were created in 12 swine. In each animal, one aneurysm was embolized using BPM/GDCs and the other aneurysm was embolized using standard GDCs. Comparative angiographic and histopathological data were analyzed at 2 weeks and 3 months postembolization. At 14 days postembolization, angiograms revealed evidence of neck neointima in six of eight aneurysms treated with BPM/GDCs compared with zero of eight aneurysms treated with standard GDCs (p < 0.05). At 3 months postembolization, angiograms demonstrated that four of four aneurysms treated with BPM/GDC were smaller and had neck neointima compared with zero of four aneurysms treated with standard GDCs (p = 0.05). At 14 days, histological analysis of aneurysm healing favored BPM/GDC treatment (all p < 0.05): the grade of cellular reaction around the coils was 3 ± 0.9 (mean ± standard deviation) for aneurysms treated using BPM/GDCs compared with 1.6 ± 0.7 for aneurysms treated using GDCs alone; the percentage of unorganized thrombus was 16 ± 12% compared with 37 ± 15%, and the neck neointima thickness was 0.65 ±0.26 mm compared with 0.24 ±0.21 mm, respectively. At 3 months postembolization, only neck neointima thickness was significantly different (p < 0.05): 0.73 ± 0.37 mm in aneurysms filled with BPM/GDCs compared with 0.16 ± 0.14 mm in aneurysms filled with standard GDCs. Conclusions. In experimental aneurysms in swine, BPM/GDCs accelerated aneurysm fibrosis and intensified neck neointima formation without causing parent artery stenosis or thrombosis. The use of BPM/GDCs may improve long-term anatomical outcomes by decreasing aneurysm recanalization due to stronger in situ anchoring of coils by organized fibrous tissue. The retraction of this scar tissue may also decrease the size of aneurysms and clinical manifestations of mass effect observed in large or giant aneurysms.Keywords
This publication has 39 references indexed in Scilit:
- Fibrinogen and Vascular Smooth Muscle Cell Grafts Promote Healing of Experimental Aneurysms Treated by EmbolizationStroke, 1999
- Ion Implantation and Protein Coating of Detachable Coils for Endovascular Treatment of Cerebral Aneurysms: Concepts and Preliminary Results in Swine ModelsNeurosurgery, 1997
- Potential efficacy of basic fibroblast growth factor incorporated in biodegradable hydrogels for skull bone regenerationJournal of Neurosurgery, 1997
- Effect of (poly)-L-lactic acid on the proliferation and differentiation of primary bone cells in vitroJournal of Biomedical Materials Research, 1996
- Microsurgically produced bifurcation aneurysms in a rabbit model for endovascular coil embolizationJournal of Neurosurgery, 1996
- Effect of Guglielmi Detachable Coils on Experimental Carotid Artery Aneurysms in PrimatesStroke, 1995
- Treatment of Experimental Aneurysms Using Collagen-coated Microcoils Experimental StudyNeurosurgery, 1995
- Treatment of Experimental Aneurysms Using Collagen-coated MicrocoilsNeurosurgery, 1995
- Tissue EngineeringScience, 1993
- New Methods of Drug DeliveryScience, 1990