A high-performance liquid chromatographic method using a microbore system consisting of a microbore column (100 mm – 1mm I. D.- 3 μm), microbore pump (60 μl/min) and UV absorbance detector has been developed for the determination of adenosine in rat brain microdialysis samples. UV absorbance of adenosine at 259 nm was used to measure adenosine. The limit of detection (LOD) was 50 pg and the limit of quantitation (LOQ) was 10 ng/ml. The methodology is applicable for the quantitation of adenosine in small volumes (3 – 10 (μl) of microdialysate and does not require sample preparation.Keywords
This publication has 4 references indexed in Scilit:
- Subtraction method for the high-performance liquid chromatographic measurement of plasma adenosineJournal of Chromatography B: Biomedical Sciences and Applications, 1992
- Quantitation of adenosine, inosine and hypoxanthine in biological samples by microbore-column isocratic high-performance liquid chromatographyJournal of Chromatography A, 1991
- Application of a Variance-Stabilizing Transformation Approach to Linear Regression of Calibration LinesJournal of Pharmaceutical Sciences, 1990
- Extraction and measurement of myocardial nucleotides, nucleosides, and purine bases by high-performance liquid chromatographyAnalytical Biochemistry, 1988