Dose rate dependence of a PTW diamond detector in the dosimetry of a 6 MV photon beam
- 1 August 1994
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 39 (8) , 1219-1229
- https://doi.org/10.1088/0031-9155/39/8/003
Abstract
The dose rate dependence and current/voltage characteristics of a PTW Riga diamond detector in the dosimetry of a 6 MV photon beam have been investigated. Diamond detectors are radiosensitive resistors whose conductivity (i) varies almost in proportion to dose rate and (ii) is almost independent of bias voltage for a constant dose rate. At the recommended bias of +100 V, and also at +200 V, the detector is operating with incomplete charge collection due to the electron-hole recombination time being shorter that the maximum time for an electron to be collected by the anode. As dose rate is varied by changing FSD or depth (changing dose per pulse), detector current and dose rate are related by the expression i alpha DDelta where Delta is approximately 0.98. This manifests itself in an overestimate in percentage depth-dose at a depth of 30 cm of approximately 1% when compared to ionization chamber results. A similar sublinearity is seen when pulse repetition frequency is varied, indicating that the dependence is an on average rather than an instantaneous dose rate. The dose rate dependence is attributed to the reduction in recombination time as dose rate increases.Keywords
This publication has 10 references indexed in Scilit:
- Evaluation of a PTW diamond detector for electron beam measurementsPhysics in Medicine & Biology, 1993
- Application of a natural diamond detector for the measurement of relative dose distributions in radiotherapyPhysics in Medicine & Biology, 1993
- Diamond Detectors in Relative Dosimetry of Photon, Electron and Proton Radiation FieldsRadiation Protection Dosimetry, 1990
- Characteristics of a P-Si Detector in High Energy Electron FieldsActa Radiologica: Oncology, 1985
- Dosimetry with a diamond operating as a resistorPhysics in Medicine & Biology, 1981
- Evaluation of diamond radiation dosemetersPhysics in Medicine & Biology, 1980
- Electrical properties and performances of natural diamond nuclear radiation detectorsNuclear Instruments and Methods, 1979
- Diamond dosimeter for x-ray and δ-radiationIEEE Transactions on Nuclear Science, 1977
- Preparation and Characteristics of Natural Diamond Nuclear Radiation DetectorsIEEE Transactions on Nuclear Science, 1975