Thermographic study of laser on arteries
- 1 January 1987
- journal article
- research article
- Published by Wiley in Lasers in Surgery and Medicine
- Vol. 7 (4) , 307-329
- https://doi.org/10.1002/lsm.1900070405
Abstract
We analyzed the different effects of CO2, Nd-YAG, and argon lasers on aorta by using a Thermovision infrared system that registered the laser interaction with mongrel dog aorta. The images (thermograms) obtained were processed by a computer, which associated each area of the same temperature with a determined color. These thermograms were compared to histological analysis of the respective samples and the following results were obtained: (1) After the application of each laser there is very little propagation of heat in tissue. (2) The CO2 laser makes tissue reach 100°C in less than 0.05 seconds. (3) The heat dissipation time was higher with the Nd-YAG laser due to higher scattering on tissue. Based on this research we conclude that the CO2 laser was best absorbed, the Nd-YAG laser penetrated human tissue with the best results, and the argon laser had the most significant backscattering.Keywords
This publication has 5 references indexed in Scilit:
- Arterial response to laser operation for removal of atherosclerotic plaquesThe Journal of Thoracic and Cardiovascular Surgery, 1983
- The Biomedical LaserPublished by Springer Nature ,1981
- Lasers in Biology and MedicinePublished by Springer Nature ,1980
- Optics and LasersPublished by Springer Nature ,1977