Breathing systems: effect of fresh gas flow rate on enflurane consumption
Open Access
- 1 December 1994
- journal article
- research article
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 73 (6) , 775-778
- https://doi.org/10.1093/bja/73.6.775
Abstract
The vaporization rates of enflurane were measured in 412 anaesthetics using appropriate fresh gas flow rates in Bain (12 litre min−1 Magill (6 litre min−1) and circle systems (3 litre min−1, 1 litre min−1 and “closed”). In all patients reducing the fresh gas flow rate resulted in lower enflurane consumption. The percent savings were 18–86% depending on the initial fresh gas flow rate and the size of the change in fresh gas flow. The reduction in enflurane use was more marked in inpatients (long cases) than in day-case patients (short cases).Keywords