A micromethod for the determination of fluoride in blood plasma and saliva
- 1 December 1977
- journal article
- research article
- Published by Springer Nature in Calcified Tissue International
- Vol. 23 (1) , 225-228
- https://doi.org/10.1007/bf02012789
Abstract
A simple and accurate technique for the determination of fluoride (F−) in capillary-sampled blood is presented. The method is based on the known addition-slope determination technique using the fluoride electrode is required. A standard deviation of 1.3–5.6% in the range 300–10 ng F/ml was given by 259 duplicate determinations on human plasma. Measurements of parotid saliva showed that it was possible to analyze fluoride concentrations as low as 4 ng F/ml with good reproducibility. The results also show that a 150 μl sample is sufficient for fluoride determination with high accuracy. This method can easily be used in a clinical laboratory.This publication has 8 references indexed in Scilit:
- Therapie der Osteoporose mit FluoridDeutsche Medizinische Wochenschrift (1946), 1976
- Fluoride determination in plasma by ion selective electrodes: A simplified method for the clinical laboratoryClinica Chimica Acta; International Journal of Clinical Chemistry, 1975
- Reverse extraction technique for the determination of fluoride in biological materialsAnalytical Chemistry, 1974
- Three Years' Experience with Sodium-Fluoride Therapy of OsteoporosisActa Orthopaedica, 1974
- Direct Potentiometric Determination of Total Ionic Fluoride in Biological FluidsClinical Chemistry, 1972
- Serum fluoride analysis with the fluoride electrode.1970
- The direct determination of fluoride in human saliva by a fluoride electrodeArchives of Oral Biology, 1968