Inoculum size and lot-to-lot variation as significant variables in the tube coagulase test for Staphylococcus aureus
- 1 November 1985
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 22 (5) , 835-837
- https://doi.org/10.1128/jcm.22.5.835-837.1985
Abstract
For a positive coagulase test with certain isolates of Staphylococcus aureus, inoculation of 0.5 ml of coagulase plasma with at least two colonies and 24 h of incubation at 35 to 37.degree. C may be required. S. aureus does not multiply in coagulase plasma. There is also lot-to-lot variation in the sensitivity of coagulase plasma. On isolates from critical specimens, a negative coagulase test obtained with an inoculum of less than two colonies should be confirmed by repeating the procedure with a larger inoculum or by using some other confirmatory test.This publication has 6 references indexed in Scilit:
- Comparison of lysis-centrifugation with lysis-filtration and a conventional unvented bottle for blood culturesJournal of Clinical Microbiology, 1984
- Evaluation of three commercial agglutination tests for the identification of Staphylococcus aureusJournal of Clinical Microbiology, 1984
- Evaluation of latex agglutination and microtube coagulase tests for detection of Staphylococcus aureusDiagnostic Microbiology and Infectious Disease, 1984
- Two coagulase-variant forms of Staphylococcus aureus isolated from blood culturesJournal of Clinical Microbiology, 1984
- Microtube coagulase test for detection of coagulase-positive staphylococciJournal of Clinical Microbiology, 1982
- Characterization and Identification of Coagulase-negative, Heat-stable Deoxyribonuclease-positive StaphylococciJournal of General Microbiology, 1978