The Accuracy of the CGMS™in Children with Type 1 Diabetes: Results of the Diabetes Research in Children Network (DirecNet) Accuracy Study
- 1 October 2003
- journal article
- clinical trial
- Published by Mary Ann Liebert Inc in Diabetes Technology & Therapeutics
- Vol. 5 (5) , 781-789
- https://doi.org/10.1089/152091503322526987
Abstract
The accuracy of the Continuous Glucose Monitoring System, CGMS™ (Medtronic MiniMed, Northridge, CA) was assessed in children and adolescents with type 1 diabetes mellitus (T1DM) when compared with reference serum glucose levels during spontaneous fluctuations in glucose levels over 24 h and during acute hyper- and hypoglycemia. Ninety-one subjects with type 1 diabetes (3.5-17.7 years old) wore one or two CGMSs while blood samples were obtained for serum glucose determinations (made at a central laboratory) hourly during the day, every 30 min overnight, and every 5 min during meal-induced hyperglycemia and insulin-induced hypoglycemia tests, resulting in 6,778 CGMS-reference glucose pairs. CGMS function was assessed on each of the 3 days of sensor life. The median relative absolute difference (RAD) between the CGMS and reference values was 18% (25th, 75th percentiles = 8%, 34%). Similar results were obtained on each of the 3 days of sensor life. Accuracy was worse during hypoglycemia than during hyperglycemia. Modified sensors that first became available in November 2002 were more accurate than were the original sensors (median RAD = 11% vs. 19%) and had better precision (r = 0.92 vs. r = 0.77) during time periods in which two CGMSs were simultaneously used. The CGMS sensors that have been in clinical use until recently are often inaccurate in quantifying glucose values in children with T1DM. However, recent modifications to the sensor have resulted in substantially better accuracy and reliability. This improved function, if confirmed by additional data, may enhance the clinical utility of the CGMS.Keywords
This publication has 16 references indexed in Scilit:
- The Accuracy of the GlucoWatch®G2™Biographer in Children with Type 1 Diabetes: Results of the Diabetes Research in Children Network (DirecNet) Accuracy StudyDiabetes Technology & Therapeutics, 2003
- Clinical Performance of CGMS in Type 1 Diabetic Patients Treated by Continuous Subcutaneous Insulin Infusion Using Insulin AnalogsDiabetes Care, 2003
- Hypoglycemia Prevalence in Prepubertal Children With Type 1 Diabetes on Standard Insulin Regimen: Use of Continuous Glucose Monitoring SystemDiabetes Care, 2003
- Nocturnal hypoglycemia detected with the continuous glucose monitoring system in pediatric patients with type 1 diabetesThe Journal of Pediatrics, 2002
- Spurious Reporting of Nocturnal Hypoglycemia by CGMS in Patients With Tightly Controlled Type 1 DiabetesDiabetes Care, 2002
- Performance Evaluation of the MiniMed® Continuous Glucose Monitoring System During Patient Home UseDiabetes Technology & Therapeutics, 2000
- Comparison of Metabolic Control in a Cross-Sectional Study of 2,873 Children and Adolescents With IDDM from 18 CountriesDiabetes Care, 1997
- Reviewing the Diabetes Control and Complications Trial: One member of the "control panel" speaksThe Journal of Pediatrics, 1994
- Effect of intensive diabetes treatment on the development and progression of long-term complications in adolescents with insulin-dependent diabetes mellitus: Diabetes Control and Complications TrialThe Journal of Pediatrics, 1994
- The Effect of Intensive Treatment of Diabetes on the Development and Progression of Long-Term Complications in Insulin-Dependent Diabetes MellitusNew England Journal of Medicine, 1993