An evaluation of CT systems from ten manufacturers
- 1 December 1981
- journal article
- research article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 54 (648) , 1053-1061
- https://doi.org/10.1259/0007-1285-54-648-1053
Abstract
During the past three years, 23 computerized tomography (CT) machines from ten different manufacturers located in Europe and the USA have been evaluated. Using the same set of phantoms and measurement techniques, performance data have been derived including noise, uniformity of CT numbers, low contrast detectability, spatial resolution, modulation transfer function, effective photon energy and linearity of CT numbers. Thermoluminescent and film dosimetry were used to determine single-slice dose profiles at the periphery and centre of specialized anthropomorphic phantoms. The results of this study are presented in tabular and graphical form together with some general conclusions on their implications.Keywords
This publication has 5 references indexed in Scilit:
- Evaluating performance characteristics in computerized tomographyThe British Journal of Radiology, 1981
- A survey of 29 EMI CT machines in BritainThe British Journal of Radiology, 1981
- The measurement of effective photon energy and “linearity” in computerized tomographyThe British Journal of Radiology, 1980
- The line spread function and modulation transfer function of a computed tomographic scannerMedical Physics, 1976
- Computerized transverse axial scanning (tomography): Part 1. Description of systemThe British Journal of Radiology, 1973