Microbial validation of vent filters
- 1 August 1984
- journal article
- Published by Wiley in Biotechnology & Bioengineering
- Vol. 26 (8) , 828-835
- https://doi.org/10.1002/bit.260260803
Abstract
The Upjohn Company uses filtration to remove microorganisms and particulates from air and other gases which may come in contact with sterile products. To validate the microbial retentivity of these filters, they were challenged with an aerosol of Bacillus subtilis var niger spores. An aerosol challenge was used because it more closely simulated the use for which these filters were designed. The test apparatus was constructed of autoclavable components using a jet‐type nebulizer and heated air mixing tube. Characterization of the aerosol particle size distribution with a particle size analyzer demonstrated that 80% of the particles had a diameter of × 3.0 times;m and that the particles had a mean mass diameter of 1.9 times;m with a geometric standard deviation of 1.8 times;m. Studies conducted with aerosols of Bacillus subtilis var niger spores demonstrated that the test apparatus could recover ca. 50% of the spores that were aerosolized. Hydrophobic filters from various manufactures were challenged with an aerosol of at least 108 spores of Bacillus subtilis. All filters tested could retain at least 109 spores when physical integrity of the filter was verfield.Keywords
This publication has 5 references indexed in Scilit:
- Homogeneous Bacterial Aerosols Produced with a Spinning-Disc GeneratorApplied Microbiology, 1970
- Homogeneous Bacterial Aerosols Produced with a Spinning-Disc GeneratorApplied Microbiology, 1970
- Sampling Submicron T1 Bacteriophage AerosolsApplied Microbiology, 1965
- Sampling Submicron T1 Bacteriophage AerosolsApplied Microbiology, 1965
- Physical Tracers for Bacterial AerosolsApplied Microbiology, 1961