Simple, reliable, and sensitive interferometer for the measurement of the refractive index of liquids as a function of temperature
- 1 April 1979
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 50 (4) , 509-510
- https://doi.org/10.1063/1.1135861
Abstract
We describe a very simple and inexpensive interferometer setup which provides high accuracy measurements of the refractive index (n) of liquids with temperature. A standard deviation sigma (n) =5.4x10(-7) has been observed in isooctane liquid when fitting n to a linear temperature variation.Keywords
This publication has 2 references indexed in Scilit: