A new device has been developed to estimate continuously and noninvasively cardiac output from the thoracic electrical bioimpedance (CObi). CObi was compared to cardiac output by thermodilution (COtd) in five anesthetized dogs. Blood pressure, blood volume, and blood flow were manipulated by hemorrhage and infusions of sodium nitroprusside and phenylephrine. These data were used to determine the correlation between CObi and COtd under conditions of hypotensive normal flow and normotensive low flow, as well as during hemorrhagic shock and resuscitation. The CObi device was calibrated in vivo to COtd for each dog at the beginning of each experiment. CObi had a significant positive correlation with COtd throughout the experiments (r = 0.84, slope = 0.91, intercept = 0.55, pKeywords