Fiber optic immunochemical sensor for continuous, reversible measurement of phenytoin.
Open Access
- 1 July 1988
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 34 (7) , 1417-1421
- https://doi.org/10.1093/clinchem/34.7.1417
Abstract
This fiber optic sensor, based on a homogeneous fluorescence energy-transfer immunoassay, operates in a continuous, reversible manner to quantify the anticonvulsant drug phenytoin. B-Phycoerythrin-phenytoin and Texas Red-labeled antibody to phenytoin were sealed inside a short length of cellulose dialysis tubing, which was cemented to the distal end of an optical fiber. When the sensor was alternately placed into solutions with various concentrations of free phenytoin, the drug crossed the dialysis membrane and displaced a fraction of the B-phycoerythrin-phenytoin from the antibody. The resulting change in fluorescence signal was measured with a fiber optic fluorometer. A typical competitive-binding calibration curve was seen between 5 and 500 mumol of phenytoin per liter. Equilibrium response time ranged from 5 to 30 min for different sensors. Replicate equilibrium measurements with one sensor, alternated eight times between two solutions of phenytoin, gave a CV of 2.1% (n = 16) at 0 mumol/L and 2.4% (n = 13) at 100 mumol/L. A reversible immunochemical sensor can be made that has a response time suitable for continuous concentration measurements.This publication has 9 references indexed in Scilit:
- Fiber-optic biosensor for ethanol, based on an internal enzyme conceptTalanta, 1988
- Polymer immobilized enzyme optrodes for the detection of penicillinAnalytical Chemistry, 1987
- Fiber-optic chemical sensors for competitive binding fluoroimmunoassayAnalytical Chemistry, 1987
- Physiological pH fiber-optic chemical sensor based on energy transferAnalytical Chemistry, 1987
- Fiber-optic probe for in vivo measurement of oxygen partial pressureAnalytical Chemistry, 1984
- A fiber optic PCO2 sensorAnnals of Biomedical Engineering, 1983
- Texas red, a hydrophilic, red-emitting flourophore for use with flourescein in dual parameter flow microfluorometric and fluorescence microscopic studiesJournal of Immunological Methods, 1982
- Fiber optic pH probe for physiological useAnalytical Chemistry, 1980
- Fluorescent excitation transfer immunoassay. A general method for determination of antigens.Journal of Biological Chemistry, 1976