Trace concentrations of mercury in water samples were determined by a method involving a preconcentration procedure and inductively coupled plasma mass spectrometry (ICP-MS) detection. Mercury vapour, generated by tin(II) chloride as the reductant, was trapped by amalgamation on a gold–platinum gauze, released by controlled heating and detected by ICP-MS. A flow injection sample introduction system with time-based injection was used and the sensitivity was found to be proportional to the mass of mercury introduced. Quantitative recoveries of mercury from open ocean sea-water, coastal sea-water and fresh water reference materials were obtained. A detection limit of 200 pg l–1 for a 25 ml sample was obtained. The precision of the method was 1%(relative standard deviation) at 1 µg l–1.Keywords
This publication has 17 references indexed in Scilit:
- On-line microwave sample pretreatment for the determination of mercury in blood by flow injection cold vapor atomic absorption spectrometryTalanta, 1993
- Determination of mercury in urine by flow-injection cold vapour atomic absorption spectrometryAnalytica Chimica Acta, 1993
- On-line preconcentration and determination of mercury by flow injection inductively coupled plasma atomic emission spectrometryJournal of Analytical Atomic Spectrometry, 1993
- Design and optimisation of a flow injection hydride generator and its use for automated standard additionsJournal of Analytical Atomic Spectrometry, 1990
- On-line separation and preconcentration in flow injection analysisAnalytica Chimica Acta, 1988
- Adaptation of a cold vapour mercury analyser to flow injection analysisJournal of Analytical Methods in Chemistry, 1988
- Determination of uranium in environmental samples using inductively coupled plasma mass spectrometryAnalytical Chemistry, 1987
- Combination of flow-injection techniques with atomic spectrometry in agricultural and environmental analysisAnalytica Chimica Acta, 1986
- Cold Vapor Atomic Absorption determination of mercury by Flow Injection Analysis using a Teflon membrane phase separator coupled to the absorption cellSpectrochimica Acta Part B: Atomic Spectroscopy, 1983
- Inductively coupled argon plasma as an ion source for mass spectrometric determination of trace elementsAnalytical Chemistry, 1980