Stability: An ICD Detection Criterion for Discriminating Atrial Fibrillation From Ventricular Tachycardia
- 1 December 1995
- journal article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 6 (12) , 1081-1088
- https://doi.org/10.1111/j.1540-8167.1995.tb00385.x
Abstract
The purpose of this study was to review a new implantable cardioverter defibrillator (ICD) detection criterion, stability, to determine if it can effectively discriminate rapid rhythms of atrial fibrillation from ventricular tachycardia. Inappropriate shocks for rapid atrial fibrillation limit the acceptance of ICDs. The advent of an additional detection criterion, stability, has been postulated to be of value in discriminating rapid atrial fibrillation, which may not warrant treatment, from ventricular tachycardia, which obviously does warrant therapy deliver. Twenty-six patients were studied during 32 episodes of rapid atrial fibrillation and 24 episodes of monomorphic ventricular tachycardia below 220 beats/min. Each rhythm was repeatedly evaluated by the device at each of the seven stability values available (8, 16, 23, 31, 39, 47, and 55 msec) and then classified as stable or unstable. Upon completion of this acute study, 32 ICD patients had the stability feature activated and were followed for proper arrhythmia treatment by the device. Using stability windows from 8 to 47 msec, all atrial fibrillation rhythms were appropriately classified as unstable. Three of 6 were classified correctly for the 55-msec window. All ventricular tachycardia rhythms were appropriately classified as stable from all stability windows from 8 to 55 msec. Clinical follow-up confirmed appropriate therapy delivery when coupled with sustained rate duration (SRD). Thirty-two patients followed for 292 patient-months had no episodes of untreated ventricular tachycardia with 428 successfully classified as stable and treated. Only three episodes of suspected atrial fibrillation resulted in therapy delivery as the rhythm duration exceeded the SRD of 30 seconds. The CPI Ventak PRx ICD is highly reliable in appropriately classifying atrial fibrillation as unstable and monomorphic ventricular tachycardia as stable for most stability windows evaluation tachycardias below 220 beats/min. As a result, when testing of atrial fibrillation is not possible, we recommend the routing programming of this stability feature at the 31-msec window with an SRD of 30 seconds. The reliability of this device in discriminating atrial fibrillation from monomorphic ventricular tachycardia may have important clinical implications for other tiered therapy ICDs with this feature as well as for future ICDs in development.Keywords
This publication has 18 references indexed in Scilit:
- Underdetection of ventricular tachycardia by algorithms to enhance specificity in a tiered-therapy cardioverter-defibrillatorJournal of the American College of Cardiology, 1994
- Clinical experience with antitachycardia pacing and improved detection algorithms in a new implantable cardioverter-defibrillatorJournal of the American College of Cardiology, 1993
- Electrocardiographically Documented Unnecessary, Spontaneous Shocks in 241 Patients With Implantable Cardioverter DefibrillatorsPacing and Clinical Electrophysiology, 1992
- The automatic implantable cardioverter defibrillator: Efficacy, complications and survival in patients with malignant ventricular arrhythmiasJournal of the American College of Cardiology, 1988
- The automatic implantable cardioverter-defibrillator: An overviewJournal of the American College of Cardiology, 1985
- Clinical experience, complications, and survival in 70 patients with the automatic implantable cardioverter/defibrillator.Circulation, 1985
- Clinical evaluation of the internal automatic cardioverter-defibrillator in survivors of sudden cardiac deathThe American Journal of Cardiology, 1983
- Torsades de pointes: electrophysiologic studies in patients without transient pharmacologic or metabolic abnormalities.Circulation, 1981
- Termination of Malignant Ventricular Arrhythmias with an Implanted Automatic Defibrillator in Human BeingsNew England Journal of Medicine, 1980
- Mechanism of ventricular fibrillation in man: Observations based on electrode catheter recordingsThe American Journal of Cardiology, 1979