Dual X-ray absorptiometry forearm software: Accuracy and intermachine relationship
- 1 September 1994
- journal article
- clinical trial
- Published by Oxford University Press (OUP) in Journal of Bone and Mineral Research
- Vol. 9 (9) , 1425-1427
- https://doi.org/10.1002/jbmr.5650090914
Abstract
An ash study was performed to analyze the accuracy of forearm measurements at the one‐third site of three dual x‐ray absorptiometry (DXA) systems using 20 cadavers. A Hologic QDR‐2000, a Hologic QDR‐1000/W, and a Lunar DPX‐L system were used. The correlations between ash weight and DXA BMC were excellent for the three instruments (r > 0.97, p < 0.001), with accuracy errors < 5.2%. To perform a forearm cross‐calibration of bone mineral content (BMC) and density (BMD) between the scanners, 10 healthy volunteers were additionally scanned at the distal one‐third radius. The correlations among the DXA machines were excellent (r > 0.95); the absolute BMC and BMD values were significantly different between the two Hologics and the Lunar machine. The slope and intercept of both the BMC and BMD between the two Hologic systems were close to unity. In conclusion, the DXA forearm software packages provide accurate methods for assessing bone mineral content and density. The conversion of data among different manufacturers should be performed by careful cross‐calibration measurements.Keywords
This publication has 12 references indexed in Scilit:
- Cross-calibration of DXA equipment: Upgrading from a hologic QDR 1000/W to a QDR 2000Calcified Tissue International, 1993
- Assessment of spinal and femoral bone density by Dual X-Ray absorptiometry: Comparison of lunar and hologic instrumentsJournal of Bone and Mineral Research, 1992
- Comparison of single-photon and dual-energy x-ray absorptiometry of the radiusBone and Mineral, 1992
- Dual-energy X-ray absorptiometry of the forearm: Reproducibility and correlation with single-photon absorptiometryJournal of Bone and Mineral Research, 1992
- A comparison of two dual-energy X-ray absorptiometry systems for spinal bone mineral measurementCalcified Tissue International, 1992
- Dual-energy X-ray absorptiometry versus single photon absorptiometry of the radiusCalcified Tissue International, 1991
- Age and bone mass as predictors of fracture in a prospective study.Journal of Clinical Investigation, 1988
- In Vivo Measurement of Bone MassJAMA, 1972
- Measurement of Bone Mineral in vivo: An Improved MethodScience, 1963