Load sharing and graft forces in anterior cruciate ligament reconstructions with the Ligament Augmentation Device
- 1 May 1989
- journal article
- research article
- Published by SAGE Publications in The American Journal of Sports Medicine
- Vol. 17 (3) , 414-422
- https://doi.org/10.1177/036354658901700317
Abstract
Buckle transducers were used to directly measure load sharing and overall force in three segment (semitendi nosus, gracilis, Ligament Augmentation Device) and two segment (patellar tendon, LAD) composite ACL grafts during the application of anteriorly directed loads to a series of five fresh knee specimens with a pneu matic load apparatus. The total graft forces generated in the two segment and three segment composite reconstructions during this immediate postoperative state were highly variable when compared to the normal ACL in each specimen, and were irreproducible among the five specimens. Load sharing among the graft com ponents occurred and was also variable in both recon structions, with this variation being greater in the three segment graft. The LAD carried an average of 45% of the total graft force in the three segment graft and 28% in the two segment graft. Further work is required to clarify the source of the variability in the total graft force and load sharing observed in this study.This publication has 19 references indexed in Scilit:
- Description and Error Evaluation of an In Vitro Knee Joint Testing SystemJournal of Biomechanical Engineering, 1988
- Polypropylene braid augmented and nonaugmented intraarticular anterior cruciate ligament reconstructionThe American Journal of Sports Medicine, 1985
- A High-strength Dacron Augmentation for Cruciate Ligament ReconstructionPublished by Wolters Kluwer Health ,1985
- On the Interpretation of Our Anterior Cruciate Ligament DataPublished by Wolters Kluwer Health ,1985
- Biomechanics of Cranial Cruciate Ligament Reconstruction in the Dog II. Mechanical PropertiesVeterinary Surgery, 1983
- Arthroscopic Repair and Augmentation of the Anterior Cruciate Ligament in Cadaver KneesPublished by Wolters Kluwer Health ,1983
- Vascularized Patellar Tendon Graft with Rigid Internal Fixation for Anterior Cruciate Ligament InsufficiencyClinical Orthopaedics and Related Research, 1983
- Intra-articular Cruciate ReconstructionPublished by Wolters Kluwer Health ,1983
- A Note on the Application and Evaluation of the Buckle Transducer for Knee Ligament Force MeasurementJournal of Biomechanical Engineering, 1982
- Presidential addressThe American Journal of Sports Medicine, 1980