Abstract
A large (22 × 25 mm) semi-transparent thin film of gold, approximately 400 Å in thickness, is deposited on a glass substrate for simultaneous use as a heat source and resistance thermometer. Construction techniques and calibration procedures are described, and a sample application to a transient boiling process is included with simultaneous high speed photographs taken through the thin film from beneath.