Determination of selenium in urine by inductively coupled plasma mass spectrometry: interferences and optimization
- 1 January 1999
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 14 (5) , 867-874
- https://doi.org/10.1039/a900940j
Abstract
The aim of this study was to develop a method for selenium determination in urine and examine the influence of sensitivity enhancement reagents, instrument parameters, internal standards and the salt content of the urine matrix on the determination. Several carbon-containing solutions (methanol, ethanol, propanol, butanol, glycerol, acetonitrile and acetic acid) were examined for their sensitivity enhancement effect. Enhancement factors up to six were obtained and were dependent on the nebulizer gas flow and rf power. There was no important difference in the enhancement effects of these solutes when the nebulizer gas flow rate was optimized for each solute. The influences of sample uptake rate, nebulizer flow rate and rf power were examined in multivariate experiments. The nebulizer gas flow rate and rf power were found to be interdependent, but the sample pump flow rate was independent of the other parameters. The sensitivities of different selenium species (selenite, selenate, selenomethionine and trimethylselenonium iodide) were equal during the experiments in different enhancement solutes and when analysed with the optimized parameter settings. The influence of the urine matrix and different salts on four selenium isotopes were examined. It was concluded that only 82 Se was usable for quantitative determination in urine as the blank at mass 82 was close to zero. The blank values at masses 76, 77 and 78 varied considerably and differently with different salts and salt concentrations. Yttrium, indium and gallium were compared for their use as internal standards. There was no difference in the efficiency of these internal standards and they were all usable. The accuracy of the method was determined using the NIST SRM 2670 Toxic Metals in Freeze-dried Urine (certified value 0.030±0.008 mg l –1 ). The result was 29.4±1.0 µg l –1 (n=8) and the precision was 3.4%. In synthetic urine diluted 1+1, the limit of detection was 0.9 µg l –1 and the limit of determination was 2.2 µg l –1 .Keywords
This publication has 19 references indexed in Scilit:
- Method optimization and quality assurance in speciation analysis using high performance liquid chromatography with detection by inductively coupled plasma mass spectrometrySpectrochimica Acta Part B: Atomic Spectroscopy, 1998
- Improvement of selenium determination in water by inductively coupled plasma mass spectrometry through use of organic compounds as matrix modifiersSpectrochimica Acta Part B: Atomic Spectroscopy, 1997
- Enzymatic Digestion for the Determination of Trace Elements in Blood Serum by Inductively Coupled Plasma Mass SpectrometryJournal of Analytical Atomic Spectrometry, 1997
- On-line removal of interferences via anion-exchange column separation for the determination of germanium, arsenic and selenium in biological samples by inductively coupled plasma mass spectrometryJournal of Analytical Atomic Spectrometry, 1996
- Optimization of flow injection hydride generation inductively coupled plasma mass spectrometry for the determination of selenium in water and serum samplesJournal of Analytical Atomic Spectrometry, 1995
- Carbon-enhanced inductively coupled plasma mass spectrometric detection of arsenic and selenium and its application to arsenic speciationJournal of Analytical Atomic Spectrometry, 1994
- Elimination of interferences in the determination of arsenic and selenium in biological samples by inductively coupled plasma mass spectrometryAnalytica Chimica Acta, 1993
- Comparison of chemical modifiers for simultaneous determination of different selenium compounds in serum and urine by Zeeman-effect electrothermal atomic absorption spectrometryJournal of Analytical Atomic Spectrometry, 1993
- Simple approach to reducing polyatomic ion interferences on arsenic and selenium in inductively coupled plasma mass spectrometryJournal of Analytical Atomic Spectrometry, 1989
- Elimination of chloride interference on the determination of selenium in serum by inductively coupled plasma mass spectrometryJournal of Analytical Atomic Spectrometry, 1988