A comparison between capillary and venous blood international normalized ratio determinations in a portable prothrombin time device
- 1 September 2000
- journal article
- research article
- Published by Wolters Kluwer Health in Blood Coagulation & Fibrinolysis
- Vol. 11 (6) , 559-562
- https://doi.org/10.1097/00001721-200009000-00008
Abstract
The CoaguChek PT system is a portable point-of-care device for prothrombin time testing that can be used with capillary and venous whole blood. This system had been calibrated by the manufacturer in terms of the international normalized ratio (INR) for monitoring oral anticoagulant therapy. The purpose of the present study was to compare capillary blood with venous blood INRs from healthy volunteers and patients treated with oral anticoagulants using the same CoaguChek PT system. Two different CoaguChek PT strip formulations [international sensitivity index (ISI), 1.5 and 1.1] were used in separate test series. The differences between capillary and venous blood INRs were statistically significant (P < 0.001) but the magnitude of the differences was small. The mean relative deviations of the INR were 5.8 and 2.8% for the strips with ISI 1.5 and 1.1, respectively. These deviations are clinically acceptable. It is concluded that capillary blood can be replaced by venous blood for the calibration of the CoaguChek PT system.Keywords
This publication has 9 references indexed in Scilit:
- Accuracy, Clinical Correlation, and Patient Acceptance of Two Handheld Prothrombin Time Monitoring Devices in the Ambulatory SettingAnnals of Pharmacotherapy, 1999
- Precision of Patients’ Measurements of the International Normalized Ratio (INR) Using a Patient Operated Whole Blood Home CoagulometerThrombosis Research, 1998
- Accuracy of a Portable International Normalization Ratio Monitor in Outpatients Receiving Long-Term Oral Anticoagulant TherapyThrombosis Research, 1998
- A comparison of artificially‐depleted, lyophilized coumarin and fresh coumarin plasmas in thromboplastin calibrationBritish Journal of Haematology, 1998
- Plasma tissue factor antigen levels in capillary whole blood and venous blood: Effect of tissue factor on prothrombin timeAmerican Journal of Hematology, 1997
- Multicenter evaluation of a new capillary blood prothrombin time monitoring systemBlood Coagulation & Fibrinolysis, 1995
- Comparing methods of measurement: why plotting difference against standard method is misleadingThe Lancet, 1995
- Oral AnticoagulantsChest, 1995
- Accuracy of laboratory and portable monitor international normalized ratio determinations. Comparison with a criterion standardArchives of internal medicine (1960), 1995