Reliability and validity of the combined heart rate and movement sensor Actiheart

Abstract
Accurate quantification of physical activity energy expenditure is a key part of the effort to understand disorders of energy metabolism. The Actiheart, a combined heart rate (HR) and movement sensor, is designed to assess physical activity in populations. Objective: To examine aspects of Actiheart reliability and validity in mechanical settings and during walking and running. Methods: In eight Actiheart units, technical reliability (coefficients of variation, CV) and validity for movement were assessed with sinusoid accelerations (0.1–20 m/s2) and for HR by simulated R-wave impulses (25–250 bpm). Agreement between Actiheart and ECG was determined during rest and treadmill locomotion (3.2–12.1 km/h). Walking and running intensity (in J/min/kg) was assessed with indirect calorimetry in 11 men and nine women (26–50 y, 20–29 kg/m2) and modelled from movement, HR, and movement+HR by multiple linear regression, adjusting for sex. Results: Median intrainstrument CV was 0.5 and 0.03% for movement and HR, respectively. Corresponding interinstrument CV values were 5.7 and 0.03% with some evidence of heteroscedasticity for movement. The linear relationship between movement and acceleration was strong (R2=0.99, P0.84, P<0.001) but significantly highest when combining HR and movement (SEE<1 MET). Conclusions: The Actiheart is technically reliable and valid. Walking and running intensity may be estimated accurately but further studies are needed to assess validity in other activities and during free-living. Sponsorship: The study received financial support from the Wellcome Trust and SB was supported by a scholarship from Unilever, UK.