Comparison of Enzymatic and Microbiological Gentamicin Assays
- 1 September 1974
- journal article
- research article
- Published by American Society for Microbiology in Antimicrobial Agents and Chemotherapy
- Vol. 6 (3) , 316-319
- https://doi.org/10.1128/aac.6.3.316
Abstract
The correlation coefficient between the rapid enzymatic and the overnight microbiological assays for 211 urine and serum specimens was 0.96. The 95% confidence limits yielded a correlation coefficient between 0.92 and 0.98. Both methods tended to underestimate the amount of a gentamicin added to urine. When only serum samples were considered, the predicted value obtained from the linear regression analysis of either method was within 0.57 μg/ml 99% of the time. This high degree of positive correlation will permit safe rapid adjustment of individualized patient gentamicin dosages.Keywords
This publication has 8 references indexed in Scilit:
- Gentamicin:Adenine Mononucleotide Transferase: Partial Purification, Characterization, and Use in the Clinical Quantitation of GentamicinThe Journal of Infectious Diseases, 1974
- Radioimmunoassay for Measurement of Gentamicin in BloodAntimicrobial Agents and Chemotherapy, 1973
- Radioimmunoassay of an Antibiotic: GentamicinNature New Biology, 1972
- A Rapid Chemical Assay for GentamicinNew England Journal of Medicine, 1972
- Four-Hour Microbiological Assay of Gentamicin in SerumAntimicrobial Agents and Chemotherapy, 1972
- Rapid microassay of gentamicin, kanamycin, neomycin, streptomycin, and vancomycin in serum or plasma.1971
- SIMPLE, RAPID METHOD FOR ASSAY OF AMINOGLYCOSIDE ANTIBIOTICSThe Lancet, 1971
- Gentamicin Dosages for Renal InsufficiencyAnnals of Internal Medicine, 1971