Ortho‐phthalaldehyde: a possible alternative to glutaraldehyde for high level disinfection
- 1 June 1999
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of Applied Microbiology
- Vol. 86 (6) , 1039-1046
- https://doi.org/10.1046/j.1365-2672.1999.00791.x
Abstract
Ortho-phthalaldehyde (OPA) was tested against a range of organisms including glutaraldehyde-resistant mycobacteria, Bacillus subtilis spores and coat-defective spores. Glutaraldehyde (GTA) and peracetic acid (PAA) were tested for comparative purposes. Both suspension and carrier tests were performed using a range of concentrations and exposure times. All three biocides were very effective (≥ 5 log reduction) against Escherichia coli, Staphylococcus aureus and Pseudomonas aeruginosa in suspension tests. OPA and GTA (PAA was not tested) were also very effective against Staph. aureus and Ps. aeruginosa in carrier tests. OPA showed good activity against the mycobacteria tested including the two GTA-resistant strains, but 0·5% w/v OPA was found not to be sporicidal. However, limited activity was found with higher concentrations and pH values. Coat-defective spores were more susceptible to OPA, suggesting that the coat may be responsible for this resistance. The findings of this study suggest that OPA is effective against GTA-resistant mycobacteria and that it is a viable alternative to GTA for high level disinfection.Keywords
This publication has 11 references indexed in Scilit:
- Glutaraldehyde‐resistant Mycobacterium chelonae from endoscope washer disinfectorsJournal of Applied Microbiology, 1997
- Efficacy of ‘Nu-Cidex’ (0·35% peracetic acid) against mycobacteria and cryptosporidiaJournal of Hospital Infection, 1995
- Evaluation of three glutaraldehyde-based disinfectants used in endoscopyJournal of Hospital Infection, 1995
- FDA Labeling Requirements for Disinfection of Endoscopes: A CounterpointInfection Control & Hospital Epidemiology, 1995
- The use of peracetic acid for the reprocessing of flexible endoscopes and rigid cystoscopes and laparoscopesJournal of Hospital Infection, 1995
- Glutaraldehyde: Current Status and UsesInfection Control & Hospital Epidemiology, 1994
- Bacterial spores and chemical sporicidal agentsClinical Microbiology Reviews, 1990
- High performance liquid chromatography of plasma aminoacids using orthophthalaldehyde derivatisationClinica Chimica Acta; International Journal of Clinical Chemistry, 1987
- Rapid separation of phosphoamino acids including the phosphohistidines by isocratic high-performance liquid chromatography of the orthophthalaldehyde derivativesAnalytical Biochemistry, 1985
- Interaction of the Bacillus subtilis spore protoplast, cortex, ion‐exchange and coatless forms with glutaraldehydeJournal of Applied Bacteriology, 1984