Evaluation of the GEM® Premier™ 4000: a compact blood gas CO-Oximeter and electrolyte analyzer for point-of-care and laboratory testing
- 1 January 2008
- journal article
- research article
- Published by Walter de Gruyter GmbH in cclm
- Vol. 46 (2) , 271-9
- https://doi.org/10.1515/cclm.2008.043
Abstract
GEM Premier 4000 (Instrumentation Laboratory), a new blood gas/CO-Oximeter/hematocrit/electrolyte/glucose/lactate analyzer was evaluated. The only reagents required were disposable cartridges (GEM Premier 4000 PAK comprise all components necessary for analysis, including quality controls). The evaluation was performed in two laboratories according to the Valtec protocol guidelines. Analytical performances (imprecision at three levels, linearity, inaccuracy by method comparison using patient samples, interferences) and instrument practicability were compared with three routinely used laboratory blood gas analyzers (ABL 835 and 725, Radiometer; OMNI S, Roche Diagnostics) and the centrifuged microhematocrit method. Within-run and between-run imprecision yielded good coefficients of variation values for all parameters. Linearity was satisfactory and within the expected ranges provided by the manufacturer. Method comparisons demonstrated close agreement (Deming's regression correlation coefficients 0.92-1.00). No interferences of lipemia, hyperbilirubinemia and hemolysis were found on CO-Oximetry parameters. Fetal hemoglobin >35% overestimated carboxyhemoglobin measurements, but the magnitude of error was reduced with the fetal hemoglobin correction analysis mode. The GEM Premier 4000 analytical performance was validated for all parameters with the accuracy and the reliability of traditional systems. This blood gas analyzer fulfills most of the requirements for both point-of-care and laboratory use.Keywords
This publication has 10 references indexed in Scilit:
- Potential Interference by Hydroxocobalamin on Cooximetry Hemoglobin Measurements During Cyanide and Smoke Inhalation TreatmentsAnnals of Emergency Medicine, 2007
- Hydroxocobalamin for severe acute cyanide poisoning by ingestion or inhalationThe American Journal of Emergency Medicine, 2007
- Hypermethemoglobinemia in a substance abuserThe American Journal of Emergency Medicine, 2005
- iQM: A New Concept for Quality ControlPoint of Care: The Journal of Near-Patient Testing & Technology, 2004
- Evaluation of the blood gas analyzer GEM(r) PREMIER(tm) 3000cclm, 2004
- Interference of Methylene Blue with CO-Oximetry of Hemoglobin DerivativesClinical Chemistry, 1997
- Analytical artifacts in hematocrit measurements by whole-blood chemistry analyzersClinical Chemistry, 1995
- Reduction of Neonatal Blood Eliminates the Error Induced by Oxygen Saturation in the Spectrophotometric Measurement of CarboxyhaemoglobinAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1988