Particle size measuring device in real time for dense particulate systems
- 15 February 1978
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 17 (4) , 631-634
- https://doi.org/10.1364/ao.17.000631
Abstract
The design and performance of a real-time particle size measuring device for dense particulate systems are reported. Based on measuring the ratio of scattered light at two angles, this device, in contrast to particle sizing instruments previously reported, allows more than one particle in the focal volume at any time and selects the scattered light pulses for processing. Size distributions have been determined with an average number of particles in the focal volume of up to 2.5. This represents a potential increase of more than an order of magnitude in particle concentration over single particle sizing devices.Keywords
This publication has 3 references indexed in Scilit:
- Real-time particle sizing: increasing the capability of the instantaneous scattered intensity-ratio techniqueApplied Optics, 1975
- Real Time Measurement of the Size Distribution Of Particulate Matter by a Light Scattering MethodJournal of the Air Pollution Control Association, 1973
- Particle Sizing by Means of the Forward Scattering LobeApplied Optics, 1966