Evaluation of a Multidimensional Solid-Phase Extraction Platform for Highly Selective On-Line Cleanup and High-Throughput LC−MS Analysis of Triazines in River Water Samples Using Molecularly Imprinted Polymers
- 28 April 2001
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 73 (11) , 2437-2444
- https://doi.org/10.1021/ac001483s
Abstract
A novel highly selective sample cleanup procedure based on the use of molecularly imprinted polymers (MIPs) as solid-phase extraction materials has been evaluated with respect to its applicability and routine use in environmental analysis. The method comprises the combination of a restricted access material (RAM) and a MIP allowing a selective sample preparation to be achieved in the on-line mode. This combination is called the size-selective sample separation and solvent switch (six-SPE). The RAM column combines size exclusion and adsorption chromatography, reducing the concentration of matrix molecules by a cutoff of 15 kDa. The MIP column selectively retains the triazine analytes whereas the residual matrix is not retained and separated completely. Thus, the automated RAM-MIP is capable of excluding all matrix and nontarget compounds. The cleaned and enriched extract is subsequently eluted to an HPLC column and analyzed by LC−MS. A complete on-line analysis cycle including multidimensional solid-phase extraction, separation, and detection takes less than 15 min. Terbuthylazine, atrazine, propazine, simazine, ametryn, prometryn, irgarol, and also the metabolites deethylatrazine and deisopropylatrazine can be determined without any matrix interferences, e.g., by humic acids. The whole setup is fully automated and may be continuously operated. Nonspecific interactions with the polymer are below 1% in all cases. The accuracy of the LC-MIP-LC-MS system was controlled using a certified reference material (Aquacheck). The applicability of the method to the cleanup of real samples was demonstrated by injection of contaminated river water samples. The stability of different polymers was tested by consecutive injections, and it was shown that the performance of the materials did not vary even after more than 300 enrichment and desorption cycles.Keywords
This publication has 19 references indexed in Scilit:
- Polymer- and template-related factors influencing the efficiency in molecularly imprinted solid-phase extractionsTrAC Trends in Analytical Chemistry, 1999
- Validation of new solid-phase extraction materials for the selective enrichment of organic contaminants from environmental samplesTrAC Trends in Analytical Chemistry, 1999
- Strategies in Using Analytical Restricted Access Media Columns for the Removal of Humic Acid Interferences in the Trace Analysis of Acidic Herbicides in Water Samples by Coupled Column Liquid Chromatography with UV DetectionAnalytical Chemistry, 1999
- New chemically modified polymeric resin for solid-phase extraction of pesticides and phenolic compounds from waterJournal of Chromatography A, 1998
- Determination of Phenoxy Acid Herbicides From Aqueous Samples by Improved Clean-up on Polymeric Pre-columns at High pHThe Analyst, 1997
- On-line solid-phase extraction in liquid chromatography using restricted access pre-columns for the analysis of s-triazines in humic-containing watersJournal of Chromatography A, 1996
- Simple removal of humic and fulvic acid interferences using polymeric sorbents for the simultaneous solid-phase extraction of polar acidic, neutral and basic pesticidesJournal of Chromatography A, 1996
- Determination of phenoxyacid herbicides in drinking water at the ppt-level by liquid chromatography and on-line selective preconcentrationScience of The Total Environment, 1993
- Electronic NosesAnalytical Chemistry, 1991
- Sample preparation by means of a supported liquid membrane for the determination of chlorophenoxyalkanoic acidsJournal of Chromatography A, 1989