A comparison of flash electroretinograms recorded from Burian Allen, JET, C-glide, gold foil, DTL and skin electrodes
- 1 January 1993
- journal article
- research article
- Published by Springer Nature in Eye
- Vol. 7 (1) , 169-171
- https://doi.org/10.1038/eye.1993.36
Abstract
Single flash scotopic and photopic electroretinograms (ERGs) were recorded from the same subjects using six types of corneal electrode, in order to assess their relative effectiveness. In addition, the ERG from a lower eyelid skin electrode was recorded to give an indication of the degree of attenuation to be expected from a skin electrode. On average, the scotopic ERG recorded from the Burian Allen electrode measured 471 µ? (100%), and relative to this the b-waves recorded using other electrodes were as follows: JET (89%), C-glide (77%), gold foil (56%), DTL (46%) and skin (12%). Under photopic conditions the order was the same and the interelectrode proportions similar. The b-wave amplitude recorded using the Burian Allen electrode was 125 µ? (100%), and with other electrodes was as follows: JET (93%), C-glide (78%), gold foil (60%), DTL (60%) and skin (14%).Keywords
This publication has 7 references indexed in Scilit:
- Electrophysiological assessment of visual pathway function in infantsEye, 1992
- The electroretinogram.Archives of Disease in Childhood, 1989
- Standard for Clinical ElectroretinographyArchives of Ophthalmology (1950), 1989
- Electrophysiologic testing techniques for childrenDocumenta Ophthalmologica, 1989
- ERG electrode in pedriatic patients: Comparison of DTL fiber, PVA-gel, and non-corneal skin electrodesDocumenta Ophthalmologica, 1989
- Electrodes for ElectroretinographyArchives of Ophthalmology (1950), 1986