Quantitative optical densitometry with scanning-laser film digitizers
- 9 August 1999
- journal article
- Published by Wiley in Medical Physics
- Vol. 26 (8) , 1721-1731
- https://doi.org/10.1118/1.598664
Abstract
A new process for eliminating two types of artifacts inherent in commercially available transmission scanning-laser film digitizers is presented. The first kind of artifact results in nonreproducible interference-pattern fluctuations as large as 7%. The second kind results in spreading of transmitted light from low-to-high optical density (OD) in regions with rapidly varying ODs, producing errors as large as 50%. These OD artifacts cause the loss of precision for films with low-OD regions (first type) and the loss of accuracy for films with regions of high-OD near high-OD gradients (second type). Test radiochromic films, produced by uniform exposure to a 6 MV photon beam and a high dose rate 192Ir brachytherapy source, along with test radiographic films were used to characterize the artifacts of a commercially available scanning-laser film digitizer. The interference-pattern artifact was eliminated by digitizing the films on a masked diffusing ground-glass scanning bed. The light-transmission artifact was eliminated through discrete-fast-Fourier-transform (DFFT) deconvolution of transmission profiles with measured digitizer line-spread functions. Obtaining precise OD distributions after the DFFT deconvolution required prior removal of the interference-pattern artifact and application of a low-pass Wiener noise filter. Light-transmission artifacts are particularly significant for applications requiring measurement of high-gradient OD distributions, such as brachytherapy or conformal photon-beam film dosimetry and quantitation of two-dimensional electrophoresis gels. Errors as large as 15%-35% occur in OD distributions representative of these applications. The data collection and correction process developed in this study successfully removes these artifacts.Keywords
This publication has 15 references indexed in Scilit:
- Quantitative evaluation of radiochromic film response for two‐dimensional dosimetryMedical Physics, 1997
- Dosimetry of a radioactive coronary balloon dilitation catheter for treatment of neointimal hyperplasiaMedical Physics, 1996
- High‐sensitivity GafChromic film dosimetry for 125I seedMedical Physics, 1994
- The use of radiochromic film in treatment verification of dynamic stereotactic radiosurgeryMedical Physics, 1994
- Calibration of ophthalmic applicators at NIST: A revised approachMedical Physics, 1991
- Photon energy dependence of the sensitivity of radiochromic film and comparison with silver halide film and LiF TLDs used for brachytherapy dosimetryMedical Physics, 1991
- Sensitometry of the response of a new radiochromic film dosimeter to gamma radiation and electron beamsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1991
- Acceptance testing and quality assurance of automated scanning film densitometers used in the dosimetry of electron and photon therapy beamsMedical Physics, 1983
- Film dosimetry of megavoltage photon beams: A practical method of isodensity‐to‐isodose curve conversionMedical Physics, 1981
- FILM DOSIMETRY OF HIGH‐ENERGY ELECTRONSAnnals of the New York Academy of Sciences, 1969