Calorimetric determination of the absorbed dose-to-water beam quality correction factor k Q for high-energy photon beams
- 1 November 1995
- journal article
- Published by Wiley in Medical Physics
- Vol. 22 (11) , 1749-1752
- https://doi.org/10.1118/1.597631
Abstract
A method has been developed for measuring photon beam quality correction factors kQ using direct transfer with a water calorimeter. kQ values were measured for beam qualities varying between 6 and 23 MV for a Capintec PR-06 and a PTW W30001 cylindrical ionization chamber. Measured values were intercompared with published sets of computed kQ values and agreement was found to be within measurement uncertainty (1%, one standard deviation).Keywords