Early failure of hydroxyapatite-coating in total knee arthroplasty:A case report
- 1 January 1994
- journal article
- case report
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 65 (2) , 212-214
- https://doi.org/10.3109/17453679408995438
Abstract
We observed early loosening of a hydroxyapatite-coated tibial component in a total knee arthroplasty in a randomized study comparing HA-coating with cement fixation. Symptoms of loosening started 8 weeks after surgery. Roentgen stereophotogrammet-ric analysis revealed stability during the initial 6 weeks followed by a pronounced subsidence up to 6 months after surgery when revision was done. At revision, the HA-coating was found to have separated from the medial part of the tibial component. The synovium was abundant with macrophages and multinucleated giant cells containing HA-crystals. Overloading of the prosthesis and suboptimal quality of the HA-coating was believed to have caused the loosening process.Keywords
This publication has 3 references indexed in Scilit:
- Osteolysis from a press-fit hydroxyapatite-coated implantThe Journal of Arthroplasty, 1993
- Hydroxyapatite and urate crystal induced cytokine release by macrophages.Annals of the Rheumatic Diseases, 1989
- Extra-abdominal desmoid tumorsActa Orthopaedica, 1986