Percutaneous transluminal valvuloplasty in calcific aortic stenosis: The double balloon technique
- 1 May 1987
- journal article
- research article
- Published by Wiley in Catheterization and Cardiovascular Diagnosis
- Vol. 13 (3) , 151-156
- https://doi.org/10.1002/ccd.1810130302
Abstract
The double balloon technique for aortic valvuloplasty using a combined brachial and femoral artery approach with readily available angioplasty equipment is described. Online hemodynamic assessment of the patient during the procedure was provided by continuous, reflective spectorphotometric monitoring of the pulmonary artery mixed venous oxygen saturation and the arterial pressure in the femoral sheath. Successful valvuloplasty was achieved in 70 patients with the transaortic peak‐to‐peak systolic gradient reduced from 77.6 ± 28 to 27.8 ± 15 mmHg (P 2 (P <0.05). No significant complications were encountered. The double balloon technique for aortic valvuloplasty is an acceptable alternative method to the single balloon technique.Keywords
This publication has 7 references indexed in Scilit:
- Assessment of left ventricular and aortic valve function after aortic balloon valvuloplasty in adult patients with critical aortic stenosis.Circulation, 1987
- Geometry of the proximal pulmonary trunk during dilation with two balloonsThe American Journal of Cardiology, 1986
- Balloon dilatation of calcific aortic stenosis in elderly patients: postmortem, intraoperative, and percutaneous valvuloplasty studies.Circulation, 1986
- PERCUTANEOUS TRANSLUMINAL VALVULOPLASTY OF ACQUIRED AORTIC STENOSIS IN ELDERLY PATIENTS: AN ALTERNATIVE TO VALVE REPLACEMENT?The Lancet, 1986
- Percutaneous balloon aortic valvuloplasty: Results in 23 patientsThe American Journal of Cardiology, 1984
- Mixed venous oxygen saturation. Its role in the assessment of the critically ill patientArchives of internal medicine (1960), 1983
- Mixed Venous Oxygen SaturationArchives of internal medicine (1960), 1983